What it is
A family of electrical safety analyzers: four AC 500 VA models covering AC and DC withstanding voltage, insulation resistance, ground bond, and continuity, plus the GPT-15012 for DC withstanding up to 12 kV.
A high-efficiency PWM amplifier keeps the high-voltage output stable against fluctuating input voltage or distorted waveforms. The series is part of the GPT-10000 family, which GW Instek describes as the first safety analyzer in the world to comply with IEC 61010-2-034.
- Five models: GPT-15001 (AC), GPT-15002 (AC/DC), GPT-15003 (AC/DC/IR), GPT-15004 (AC/DC/IR/GB), GPT-15012 (DC 12 kV/IR)
- High DC withstanding (GPT-15012): 0.050 … 12.00 kV, 120 W (12 kV/10 mA)
- AC withstanding: 0.050 … 5.000 kV, 500 VA (5 kV/100 mA)
- DC withstanding: 0.050 … 6.000 kV, 100 W (5 kV/20 mA)
- Insulation resistance: 50 … 5,000 V DC, up to 50 GΩ display
- Ground bond (GPT-15004): 3.00 … 32.00 A AC, four-terminal method
- Continuity test: 100 mA DC fixed (GPT-15001 … GPT-15004)
- Measurement resolution: AC withstanding 1 µA, DC withstanding 0.1 µA, continuity 0.01 Ω
- 7-inch color TFT LCD
- Memory: 100 MANU blocks, 100 AUTO blocks
- Interfaces: RS-232C, USB device, USB host, signal I/O, remote terminal standard; GPIB or LAN optional
What it is used for
Some of the use cases for the GPT-15000 Series include:
- Safety testing to IEC, EN, UL, CSA, GB, and JIS requirements — for electronic and electrical products or parts
- Large-capacitance DUTs — such as surge absorbers
- Production line safety testing — with automatic test sequences, etc.
What does the GPT-15000 Series do?
GW Instek introduces the flagship model (500 VA output capacity) safety analyzer, the GPT-15000 series, part of the GPT-10000 family that GW Instek describes as the first safety analyzer in the world to comply with IEC 61010-2-034 (Safety requirements for electrical equipment for measurement, control, and laboratory use — particular requirements for measurement equipment for insulation resistance and test equipment for electric strength). That standard stipulates requirements for the software and hardware interfaces of high voltage and insulation resistance test and measurement instruments, so that users are provided with necessary protection and warning while using them.
The GPT-15000 models cover AC withstanding voltage, DC withstanding voltage, insulation resistance, ground bond, and continuity tests depending on model. A high-efficiency PWM amplifier excludes the influence of fluctuating input voltage or distorted waveforms, guaranteeing a stable high-voltage output during AC withstanding tests to meet safety regulations such as IEC, EN, UL, CSA, GB, and JIS.
To comply with IEC 61010-2-034, the series uses a double insulation design for input power and output voltage. A retracted on-off switch (START key) and optional activation mechanisms, for example press and hold for 1 second or pressing two keys, prevent accidental high voltage or large current output. High-illumination LED lights and a high-volume audible indicator warn of test status and judgment results, and the DUT is automatically discharged to a safe voltage of approximately 30 V after each test.
Analyze Your Safety Tests
- Meets IEC 61010-2-034 design requirements
- High accuracy and high resolution testing performance
- Flexible supplementary testing mechanism
- Statistics and analysis
- Sweep and tabular automatic test
- Complete test data retrieval interface
Why engineers pick it
- Operator protection built into the design. Double insulation on input power and output voltage, a retracted START key with optional press-and-hold or two-key activation, high-illumination LED warning lights, and a high-volume audible indicator.
- Stable output under poor mains conditions. The high-efficiency PWM amplifier excludes the influence of fluctuating input voltage or distorted waveforms during AC withstanding tests.
- Fine current resolution. AC withstanding resolves to 1 µA and DC withstanding to 0.1 µA, with ARC detection and window comparator judgment on every test.
- Automatic discharge. The DUT is discharged to approximately 30 V after each test, which matters on large-capacitance parts such as surge absorbers.
Which GPT-15000 model do I need?
| Model | AC withstanding | DC withstanding | Insulation resistance | Ground bond | Ground continuity |
| GPT-15001 | Yes | — | — | — | Yes |
| GPT-15002 | Yes | Yes | — | — | Yes |
| GPT-15003 | Yes | Yes | Yes | — | Yes |
| GPT-15004 | Yes | Yes | Yes | Yes | Yes |
| GPT-15012 | — | Yes (up to 12 kV) | Yes | — | — |
AC withstanding on the GPT-15001 through GPT-15004 has a short current above 200 mA. All models are available with optional GPIB or LAN.
When is the GPT-15000 the wrong choice?
- AC withstanding above 5 kV. AC output tops out at 5.000 kV. The 12 kV GPT-15012 is DC only and has no AC withstanding test.
- Ground bond on any model but the GPT-15004. Only the GPT-15004 provides the 3 … 32 A AC ground bond test.
- Continuity on the GPT-15012. The continuity test is available on the GPT-15001 through GPT-15004 only.
- Very high insulation resistance in ground mode. With ground mode on, the 10 … 50 GΩ range is limited to 30 GΩ maximum, adding 10% error.
Included Accessories
- All models: power cord × 1, interlock key × 1, remote terminal cable GHT-119 × 1
- GPT-15001, GPT-15002, GPT-15003: test lead GHT-115 × 1
- GPT-15004: test lead GHT-115 × 1, GTL-215 × 1
- GPT-15012: test lead GHT-120 × 1
Option
- GPT-10KG1 GPIB card
- GPT-10KL1 LAN card
Configure and order the GPT-15000
Choose the GPT-15000 model, interface card, and accessories your test station needs and order directly on this page. For questions, contact us.