What it is
An AC and DC electronic load for testing UPS, inverters, PV inverters, fuses, circuit breakers, and AC power supplies, in twenty models from 1,875 W to 22,500 W at 350 V, 425 V, and 480 V.
GW Instek launches 20 models of the AEL-5000 series AC/DC electronic loads depending on the power range. The power range of a single unit is from 1,875 W to 22,500 W, and up to 8 units can be connected in parallel. The maximum power of single-phase parallel connection can reach 180 kW, and the total power of 3-phase can reach 540 kW, which are suitable for UPS, inverter/breaker, AC power source, battery, fuse/breaker, DC power source, and other applications.
- Twenty models: 1,875 W to 22,500 W, at 350 V, 425 V, and 480 V
- Turbo mode (multiplier mode): up to 2 times the rated current and power for a short period
- Operating modes: CC, linear CC, CR, CV, CP, and AC rectifier loads
- Parallel operation: eight units (1 master, up to 7 slaves), to 180 kW single-phase and 540 kW three-phase
- Loading and unloading angle control: full range of 0 to 359 degrees
- Positive or negative half cycle load, and SCR/TRIAC phase modulation (90° trailing and leading edge)
- Inrush current and hot plug-in surge current testing
- Crest factor range: 1.414 to 5.0 (CC and CP mode)
- Frequency range: DC, 40 Hz to 440 Hz (AEL-5003-480-18.75 and AEL-5004-480-28: DC, 40 Hz to 70 Hz)
- Optional control interfaces: GPIB, RS-232, USB, and LAN
What it is used for
Some of the use cases for the AEL-5000 Series include:
- UPS testing — efficiency, back-up time, and transfer time
- PV inverter efficiency testing
- Battery discharge time testing
- Fuse and breaker trip/non-trip testing
- Short circuit, OCP, and OPP testing of AC power supplies
- Inverter startup and transient response verification, etc.
What does the AEL-5000 Series do?
The AEL-5000 series has built-in precision measurement circuits such as 16-bit A/D and DSP to provide accurate measurement items, which include voltage root mean square value (Vrms), current root mean square value (Arms), watt value (W), volt-ampere (VA), crest factor (CF), power factor (PF), total harmonic distortion (THD), voltage total harmonic distortion (VTHD), current total harmonic distortion (ITHD), peak current (Ipeak), maximum current (Amax), minimum current (Amin), maximum voltage (Vmax), minimum voltage (Vmin), and time measurement. In addition, built-in test modes include UPS efficiency, PV inverter efficiency, UPS back-up time, battery discharge time, UPS transfer time, fuse/breaker trip/non-trip, short circuit simulation, OCP, OPP, and other test modes.
The AEL-5000 series has a Turbo mode design (ON or OFF can be selected), which can increase the current and power of the electronic load by 2 times in one second. For test applications that require transient high power and large current, such as transient overload tests of protective components, short circuit of fuses and breakers, and AC power supply OCP and OPP tests, Turbo mode provides the most economical solution.
The AEL-5000 series also supports the Load On startup function (pre-set Load On). When the inverter or uninterruptible power supply is turned on, the series directly loads the set load current to verify whether startup of the inverter or uninterruptible power supply connecting to the electrical appliance is stable. At the same time, the Load On start function can also set a positive half cycle or negative half cycle load, to verify whether the output voltage of the inverter or uninterruptible power supply remains stable when the actual electrical appliance only has a positive half cycle or negative half cycle load current. Control load angle and unload angle can also be set (range 0 to 359 degrees) to verify the stability of the transient response of the inverter or uninterruptible power supply when the appliance is plugged in and unplugged. In addition, the series also supports SCR/TRIAC current phase modulation waveforms, with 90 degree trailing edge and leading edge settings.
For the adjustable bandwidth (BW) function, when the bandwidth of the DUT does not match the bandwidth of the AEL-5000 series, oscillations can occur. Users can reduce the BW setting value accordingly to meet the response speed of the DUT. Inrush current testing verifies whether the transient response of the inverter output voltage is stable when the electrical appliance is turned on (inrush current) and when the electrical appliance is suddenly connected (surge current).
The entire AEL-5000 series provides over-voltage warning, over-current, over-power, and over-temperature protection. The analog input terminal can control constant current, constant power, and other working modes through external voltage. The Vmonitor/Imonitor terminal connects an external voltage/current monitoring device. In addition, a variety of optional control interfaces are provided, such as GPIB, RS-232, USB, and LAN, to meet the needs of system integration.


Why engineers pick it
- Overload capability without a bigger load. Turbo mode doubles rated current and power for a short period, which covers fuse and breaker trip tests and AC supply short circuit, OCP, and OPP tests on a smaller unit.
- Loads the way real appliances do. Load on/off angle across 0 to 359 degrees, positive or negative half cycle loading, and SCR/TRIAC phase modulation reproduce the awkward load behavior that destabilizes inverter output.
- Test modes built in. UPS efficiency, PV inverter efficiency, UPS back-up time, battery discharge, UPS transfer time, fuse trip, and short circuit/OCP/OPP are preset functions rather than manual setups.
- Scales to 540 kW. One master and up to seven slave units reach 180 kW single-phase or 540 kW three-phase.
Which AEL-5000 model do I need?
| Power | 350 V | 425 V | 480 V |
| 1,875 W | AEL-5002-350-18.75 | AEL-5002-425-18.75 | — |
| 2,800 W | AEL-5003-350-28 | AEL-5003-425-28 | AEL-5003-480-18.75 |
| 3,750 W | AEL-5004-350-37.5 | AEL-5004-425-37.5 | AEL-5004-480-28 |
| 5,600 W | AEL-5006-350-56 | AEL-5006-425-56 | — |
| 7,500 W | AEL-5008-350-75 | AEL-5008-425-75 | — |
| 11,250 W | AEL-5012-350-112.5 | AEL-5012-425-112.5 | — |
| 15,000 W | AEL-5015-350-112.5 | AEL-5015-425-112.5 | — |
| 18,750 W | AEL-5019-350-112.5 | AEL-5019-425-112.5 | — |
| 22,500 W | AEL-5023-350-112.5 | AEL-5023-425-112.5 | — |
The 350 V models measure 50 Vrms to 350 Vrms or 500 Vdc; the 425 V models, 50 Vrms to 425 Vrms or 600 Vdc. Choose the voltage class that covers the DUT's output, then the power rating, remembering that Turbo mode provides short-period headroom above the continuous rating.
What test functions are built in?
| Function | Description |
| UPS efficiency measurement | Non-linear mode; operating frequency auto, 40 Hz to 440 Hz |
| PV systems / power conditioner efficiency (THD 80%) | Resistive + non-linear mode; operating frequency auto, 40 Hz to 440 Hz |
| UPS back-up function | CC, LIN, CR, CP |
| Battery discharge function | CC, LIN, CR, CP |
| UPS transfer time | Transfer time of off-line UPS |
| Fuse test mode | Trip/non-trip testing, with Turbo mode |
| Short / OPP / OCP test function | With Turbo mode |
| Programmable inrush current simulation | Istart – Istop / Tstep |
| Programmable surge current simulation | S1/T1 – S2/T2 – S3/T3 |
Per-model setting ranges for these test functions are listed in the manufacturer brochure.
When is the AEL-5000 the wrong choice?
- Frequencies above 440 Hz. The series covers DC and 40 Hz to 440 Hz, and the two 480 V models only reach 70 Hz.
- Below 50 Vrms. Operating voltage starts at 50 Vrms on every model.
- Sustained overload. Turbo mode doubles current and power for one second, not continuously.
- Hot environments. Operating temperature is 32 … 104 °F (0 … 40 °C).
- Bench installations on the large models. The 18,750 W and 22,500 W units stand over 50 inches tall and weigh 573 to 650 lb.